Operator’s Manual
pg. 22
5. Press the tab and lift the Fuse Holder up.
6. Replace the fuses and close the Fuse Holder completely.
7. Close the Top Cover and secure it with the two screws that you previously removed.
Cleaning/Checking the Water Filter
To clean/check the Water Filter in the ANKOM
HCl
Hydrolysis System, follow the procedure detailed below.
1. Unscrew the Water Supply Connection and barb fitting from the HCl instrument.
2. Use compressed air to clean the filter.
3. Reassemble using Teflon tape on the threads.
Tab to open
Fuse Holders
Fuse
Note the arrow
orientation
Fuse
Holders